The University of Oslo is going to acquire a Time-of-Flight Mass Spectrometer (TOF-MS) with the following specifications: compact size (flight tube length < 800 mm, reflectron geometry); bipolar ion detection (+/- switchable); mass resolving power > 5000; quadrupole ion guide interface to a vacuum chamber at 2-50 mbar to be included; analog-to-digital converter (ADC) to be included; PC with preinstalled software and hardware to be included; all vacuum pumps (turbo pump(s), prevacuum pumps; no oil pumps) and pressure gauges to be included; all to be mounted in a portable rack.
Section V: Award of contract
Contract No: 1
Lot No: 1 - Lot title: Masspectrometer TOF
V.1) Date of contract award decision:
23.9.2014
V.2) Information about offers
Number of offers received: 1
V.3) Name and address of economic operator in favour of whom the contract award decision has been taken
Tofwerk AG, Uttingenstrasse 22, 3600 Thun, SWITZERLAND